Prepare a gain reference
- First set up even illumination on the detector
- Ensure that the transmission electron microscope (TEM) is in Imaging mode, with no sample features in the spectrometer field of view
- For dedicated spectrometers
- Go to the EELS Acquisition floating window
- Set the dispersion to 0.0 eV/ch
- Start energy to 0 eV
- Set the aperture size to the minimum size
- Open the EELS Acquisition Setup Options dialog and choose a camera mode with 1x1 binning and Show 2D image checked.
- Select View in the EELS Acquisition dialog, with exposure set to 0.1 s; the intensity in the spectrum 2D view should be centered, if not adjust the Start Energy until the beam is centered
- With the spectrum view still running, increase the entrance aperture until the whole detector is completely illuminated
- Adjust the beam on the TEM to attain a count level of about 1/3 the saturation limit (21,000 for 16-bit detectors), and leave the system viewing for about 2-5 minutes to "anneal" the scintillator
- Stop the spectrum view
- Go to the EELS Acquisition floating window
- For imaging filters
- Ensure the spectrometer is in unfiltered imaging mode
- Start a camera preview with 4x binning, 0.125 s for Quantum series (HQ mode, 0.0160 s for Continuum systems) and adjust the beam to ensure even illumination on the detector.
- The image intensity should be approximately 20,000 counts (Quantum) or 350 counts (Continuum) at this binning and exposure. If not, try changing the spot size or emitter settings
- Stop the camera preview
- In the Camera menu, select Prepare Gain Reference to make sure the Target Intensity and Frames to Average are set to
- GIF 2000 and earlier Gatan imaging filter (GIF) users – 8000 and 10 respectively
- Enfina, GIF Tridiem®, GIF Quantum®, and Enfinum™ users – 21000 and 10 respectively
- Continuum™, Continuum S users – default dose rate, typical operating exposure and gain reference exposure factor are recommended
- Target does rate for gain reference is 250 – 351
- Follow any on-screen prompts during the automated gain reference acquisition procedure
Resource category
Technique category