The 3-window technique for elemental mapping requires three energy-filtered images; two positioned before the ionization edge (pre-edge images), and one positioned just after the edge.
To acquire an elemental map, specify:
Element of interest
Camera parameters
Slit width and filter energies to be used for the pre- and post-edge acquisitions
Access the Configure Elemental Map dialog will vary based on the mode you choose.
SingleMap mode – Press the Select Elemental Map button
MultiMap mode – Specify via the MultiMap Configuration dialog
Indicate parameters within the Configure Elemental Map dialog.
Window Settings (eV)
Edge – Specify the edge you want to use for ratio mapping
Pre-edge 1 and 2 – Indicate the energy loss setting that the filter will use to acquire the pre-edge image
Post-edge – Designate the energy loss setting that the filter will use to acquire the post-edge image
Slit width – Specify the energy-selecting slit width you want to use for both the pre- and post-edge acquisitions
Setup – Allows you to recall pre-saved or default settings for a particular element
Detector – Specify the detector settings you want to use
Spectrum display – Generates and displays a simulated spectrum at the top of the dialog to give you visual feedback regarding the feature shape and the selected acquisition window positions
Click Capture.
The routine will
Acquire a post-edge image followed by the two pre-edge images
Correct for any spatial drift between the images, if specified
Compute and display the elemental map